General
| Detection modes | Absorbance |
| Read methods | End point, kinetic, linear scanning |
| Microplate types | 96-well plates |
| Temperature control |
4-Zone incubation to 50 °C
+0.2 °C at 37 °C |
| Shaking | Linear |
| Software |
Gen5 Reader Control Software Gen5 Secure for 21 CFR Part 11 compliance (option) |
Absorbance
| Light source | Tungsten halogen |
| Detector | photodiode |
| Wavelength selection | filters |
| Wavelength range |
380 to 900 nm 340 to 900 nm (ELx808IU) |
| Dynamic range | 0 to 4.0 OD |
| Resolution | 0.001 OD |
| Pathlength correction | no |
| Filter wheel capacity | 6 positions |
| Filters supplied | 4 filters (5 with UV option) |
| OD accuracy | < 1% at 2.5 OD |
| OD linearity |
<1% at 2.5 OD < 2% at 3.5 OD |
| OD repeatability |
< 0.5% at 2.5 OD < 1.5% at 3.5 OD |
| Reading speed (kinetic) | 96 wells: 8 seconds |
Physical Characteristics
| Power | 100 - 240 Volts AC. 50/60 Hz |
| Dimensions |
15.5"W x 16"D x 8.75"H |
| Weight | 30 lbs (13.6 kg) |
Regulatory
| Regulatory | CE and TUV marked. RoHS Compliant. Models for In Vitro Diagnostic use are available. |