Leitor Multidetecção Híbrido Synergy Neo2
Detalhes Técnicos
General
| Detection modes |
UV-Vis absorbance Fluorescence intensity Luminescence Fluorescence polarization Time-resolved fluorescence Alpha |
| Read methods | Endpoint, kinetic, spectral scanning, well area scanning |
| Microplate types | 6- to 1536-well plates |
| Other labware supported |
Petri and cell culture dishes Take3 Micro-Volume Plates |
| Temperature control |
4-Zone incubation to 65 °C with Condensation Control™ +0.2 °C at 37 °C |
| Shaking | Linear, orbital, double orbital |
| Software |
Gen5™ Microplate Reader and Imager Software Gen5 Secure for 21 CFR Part 11 compliance (option) |
| Automation |
BioStack and 3rd party automation compatible BioSpa 8 Automated Incubator compatible |
| CO2 and O2 control (option) |
Range: 0 - 20% (CO2); 1 - 19% (O2) Control Resolution: + 0.1% (CO2 and O2) Stability: +0.2% at 5% CO2; +0.2%at 1% O2 Models for CO2 /O2 or CO2 only are available |
| Barcode reader | 1D and 2D camera-based scanner |
Absorbance
| Light source | Xenon flash |
| Detector | photodiode |
| Wavelength selection | monochromator |
| Wavelength range | 230 - 999 nm, 1 nm increments |
| Monochromator bandwidth | <3 nm (230 - 285 nm), <6 nm (>285 nm) |
| Dynamic range | 0 - 4.0 OD |
| Resolution | 0.0001 OD |
| Pathlength correction | yes |
| Monochromator wavelength accuracy | +2 nm |
| Monochromator wavelength repeatability | +0.2 nm |
| OD accuracy |
<1% at 2.0 OD <3% at 3.0 OD |
| OD linearity | <1% from 0 to 3.0 OD |
| OD repeatability | <0.5% at 2.0 OD |
| Stray light | 0.03% at 230 nm |
| Reading speed (kinetic) | 96 well: 6 seconds |
Fluorescence Intensity
| Light source | Xenon flash |
| Detector |
Dual top PMTs Single top PMT (option) Low noise PMT (bottom filter system) Red shifted PMT (top/bottom monochromator system) |
| Wavelength selection |
Quad monochromators (top/bottom) Filters (top/bottom) |
| Wavelength range |
Monochromators: 250 - 850 nm Filters (dual PMT): 200 - 850 nm |
| Monochromator bandwidth | Variable, from 3 to 50 nm in 1 nm increments |
| Dynamic range | 7 decades |
| Sensitivity |
Filters: Fluorescein 0.2 pM (4 amol/well, 20 µL, 384-well low volume plate) - top Fluorescein 1 pM (10 amol/well, 10 µL, 1536-well plate) - top Fluorescein 1 pM (0.1 fmol/well, 100 µL, 384-well plate) - bottom Quad monochromator: Fluorescein 2 pM (40 amol/well, 20 µL, 384-well low volume plate) - top Fluorescein 2.5 pM (0.25 fmol/well, 100 µL, 384-well plate) - bottom |
| Reading speed (kinetic) |
96 well: 6 seconds 384 well: 11 seconds 1536 well: 25 seconds |
Luminescence
| Wavelength range | 300 - 700 nm |
| Dynamic range | > 6 decades |
| Sensitivity |
5 amol ATP typical (384-well low volume plate) |
Fluorescence Polarization
| Light source | Xenon flash |
| Detector | Dual PMT or single PMT (option) |
| Wavelength selection | Filters |
| Wavelength range | 280 - 850 nm |
| Sensitivity | 1 mP standard deviation at 1 nM fluorescein |
Time-Resolved Fluorescence
| Light source | Xenon flash or TRF laser (option) |
| Detector | Dual PMT or single PMT (option) |
| Wavelength selection | Filters (top/bottom) |
| Wavelength range |
Monos: 250 - 850 nm
Filters (dual PMT): 200 - 850 nm |
| Sensitivity |
With TRF laser: 5 fM (384-well low volume plate) With Xenon flash lamp: 40 fM (384-well low volume plate) |
Alpha Detection
| Light source | 100 mW 680 nm laser |
| Detector | PMT |
| Wavelength selection | Filters (top) |
| Sensitivity | 100 amol LCK peptide (384-well low volume plate) |
| Read speed |
96 well: 30 seconds 384 well: 1 minute 50 seconds 1536 well: 7 minutes 20 seconds |
Reagent Dispensers
| Supported detection modes | All modes |
| Number | 2 syringe pumps |
| Supported labware | 6- to 384-well plates, Petri dishes |
| Dead volume | 1.1 mL with back flush |
| Dispense volume | 5 - 1000 µL in 1 µL increments |
| Dispense accuracy | + 1 µL or 2% |
| Dispense precision | < 2% at 50-200 µL |
Physical Characteristics
| Power | 250 Watts maximum consumption. |
| Dimensions |
Without TRF laser: 15.4" W x 20.7" D x 16.1" H (39.2 x 52.5 x 40.8 cm) With TRF laser: 15.4" W x 24.2" D x 16.1" H (39.2 x 61.4 x 40.8 cm) |
| Weight |
Without TRF laser: 78 lbs. (35.4 kg) With TRF laser: 100 lbs. (45.3 kg) |
Regulatory
| Regulatory | CE and TUV marked. RoHS Compliant. Models for In Vitro Diagnostic use are available. |