General
| Microplate types |
96- and 384-well Low profile and standard height Solid and filter bottom (filter pore sizes 0.45 µm to 1.2 µm) |
| Onboard software | Create, edit or run multiple protocols |
| Software (computer control) |
LHC2 pc software LHC2 Secure for 21 CFR Part 11 compliance (option) SiLA Compliant driver (option) |
| Separation | Biomagnetic separation, vacuum filtration (optional) |
| Shaking |
Programmable up to 60 minutes Slow, medium, fast or variable |
| Soaking | Programmable up to 60 minutes |
| Ultrasonic Advantage™ | Yes (option) |
| Verify™ clog detection | Yes (option) |
| Automation |
BioStack and 3rd party automation compatible BioSpa 8 Automated Incubator compatible |
Washing
| Manifold types |
96-well washing: 96-tube manifold - straight dispense tubes 96- and 384-well washing: 96-tube Dual-Action manifold - 20° angled dispense tubes 384-well washing (fast): 192-tube Dual-Action manifold - 7° angled dispense tubes |
| Volume range | 25 to 3,000 µL/well, in 1 µL increments |
| Wash cycles | 1 - 250 |
| Buffer/reagent selection | Auto switching (internal) for up to 4 buffers (option) |
| Supply bottle | 4 L or 10 L (optional) |
| Dispense precision |
<3% CV: 300 µL/well (96-well washing) <4% CV: 80 µL/well (384-well washing) |
| Residual volume |
<2 µL/well (96- & 384-well plates) 96-tube manifold for 96 wells; 192-tube for 384 wells |
| Wash speed |
96-wells, 300 µL/well, 3 cycles; <30 seconds 384-wells,100 µL/well, 3 cycles: <80 seconds 384-wells, 400 µL/well, 1 cycle: <20 seconds |
| Flow rates |
High flow to low flow Optimized rates for cell assays |
| Sterilization | Chemical |
| Vacuum filtration |
Selectable: Approximate final at 30 seconds: Lowest: -38 mmHg Low: -113 mmHg Medium: -200 mmHg High: -390 mmHg Highest: -506 mmHg Vacuum filtration time range: 5 to 999 seconds |
Physical Characteristics
| Power | 100 - 240 Volts AC. 50/60 Hz |
| Dimensions |
14" W x 17" D x 10" H (35.6 x 43.2 x 25.4 cm) |
| Weight |
With internal buffer switching - 36 lbs (16.5 kg) Without - 30 lbs (13.5 kg) |
Regulatory