Espectrofotômetro de Microplacas Epoch
Detalhes Técnicos
General
| Detection modes | Absorbance |
| Read methods | Endpoint, kinetic, spectral scanning, well area scanning |
| Microplate types | 6- to 384-well plates |
| Other labware supported | Take 3 Micro-Volume Plates |
| Temperature control | not available |
| Shaking | not available |
| Software |
Gen5™ Data Analysis Software Gen5 Secure for 21 CFR Part 11 compliance (option) |
| Automation | BioStack™ and 3rd party automation compatible ('R' configuration) |
Absorbance
| Light source | Xenon flash |
| Detector | photodiode |
| Wavelength selection | monochromator |
| Wavelength range | 200 to 999 nm, in 1 nm increments |
| Monochromator bandwidth | 5 nm |
| Dynamic range | 0 - 4.0 OD |
| Resolution | 0.0001 OD |
| Pathlength correction | yes |
| Monochromator wavelength accuracy | + 2 nm |
| Monochromator wavelength repeatability | + 0.2 nm |
| OD accuracy |
0 to 2 OD: +1% +0.010 OD 2 to 2.5 OD: +3% +0.010 OD |
| OD linearity |
0 to 2 OD: +% +0.010 OD 2 to 2.5 OD: +3% +0.010 OD |
| OD repeatability |
0 to 2 OD: +1% +0.005 OD 2 to 2.5 OD: +3% +0.005 OD |
| Reading speed (kinetic) |
96 wells: 15 seconds 384 wells: 31 seconds |
Physical Characteristics
| Power | 100 - 240 Volts AC. 50/60 Hz. |
| Dimensions |
12" W x 12.5" D x 7.7" H (30.5 cm x 31.8 cm x 19.6 cm) |
| Weight | <15 lbs (6.8 kg) |
Regulatory
| Regulatory | CE and TUV marked. ROHS compliant. In Vitro Diagnostic use models are available. |