Lavadora/Dispensador EL406
Detalhes Técnicos
General
| Microplate types |
96-, 384- and 1536-well Low profile and standard height Solid and filter bottom (option) |
| Onboard software | Create, edit or run multiple protocols |
| Software (computer control) |
LHC2 Software LHC2 Secure for 21 CFR Part 11 compliance (option) SiLA Compliant driver (option) |
| Separation | Biomagnetic separation, vacuum filtration (optional) |
| Shaking |
Programmable up to 60 minutes Slow, medium, fast or variable |
| Soaking | Programmable up to 60 minutes |
| Ultrasonic Advantage™ | Standard on most configurations |
| Automation |
BioStack and 3rd party automation compatible BioSpa 8 Automated Incubator compatible |
Washing
| Manifold types |
96-well washing: 96-tube manifold - straight dispense tubes 96- and 384-well washing: 96-tube Dual-Action manifold - 20° angled dispense tubes 384-well washing (fast): 192-tube Dual-Action manifold - 7° angled dispense tubes 1536-well washing: Two 32-tube dispense manifolds - 7° angled dispense tubes |
| Volume range | 3 - 3,000 µL/well in 1 µL increments |
| Wash cycles | 1 - 250 |
| Buffer/reagent selection | Auto switching module for up to 4 buffers (option) |
| Supply bottle | 4 L or 10 L (optional) |
| Dispense accuracy | +3% |
| Dispense precision | <3% CV (model dependent) |
| Residual volume | <2 µL/well |
| Wash speed |
96 wells, 300 µL/well, 96-tube manifold: 13 seconds 384 wells, 100 µL/well, 192-tube manifold: 17 seconds 1536 wells, 10 µL/wells, two 32-tube manifolds: 36 seconds |
| Flow rates |
High flow to low flow Optimized rates for cell assays |
| Sterilization | Chemical |
| Vacuum range for filtration |
Selectable: 0 to -380 mmHg (final at 30s) Filtration time: 5 - 999 seconds |
Dispensing - Peristaltic pump (Multi-Channel)
| Manifold types | 8-tip (1 x 8) cassette with plastic, 316 SS tubes or sapphire jeweled 316 SS tubes |
| Fluid delivery | 1 peristaltic pump |
| Dispense speed |
96 wells, 10 µL /well: 8 seconds 384 wells, 5 µL /well: 12 seconds 1536 wells, 1 µL /well: 27 seconds |
| Volume range | 500 nL - 3,000 µL/well, selectable in 1 µL increments |
| Flow rates |
User programmable rates from high to low Optimized rates for cell assays |
| Dispense performance |
1 µL cassette: Recommended Volume Range: 1 - 50 µL Dispense Accuracy: +5% at 1 µL Dispense Precision: <5% CV at 1 µL <10% CV at 500 nL Minimum Prime Volume: 1.20 mL 5 µL cassette: Recommended Volume Range: 5 - 2,500 µL Dispense Accuracy: +2.0% at 5 µL Dispense Precision: <2.5% CV at 5 µL Minimum Prime Volume: 4.23 mL 10 µL cassette: Recommended Volume Range: 10 - 3,000 µL Dispense Accuracy: +2.0% at 10 µL Dispense Precision: <2.0% CV at 10 µL Minimum Prime Volume: 7.36 mL |
| Recommended cassette replacement interval |
1 µL Cassette: 1,000 384-well microplates at 5 µL well 5 µL Cassette: 1,000 96-well microplates at 50 µL well 10 µL Cassette: 1,000 96-well microplates at 100 µL well |
| Sterilization | Autoclave, chemical |
Dispensing - Syringe pump
| Manifold types |
96-well dispensing: One 16-tube (2 x 8) manifold - 316 SS tubes 96-/384-well dispensing: Two 16-tube (1 x 16) manifolds - 316 SS tubes 1536-well dispensing: Two 32-tube (1 x 32) manifolds - sapphire jeweled 316 SS or 316 stainless steel tubes |
| Fluid delivery | Two positive displacement syringe drives |
| Dispense speed |
20 µL /well, 96 wells, 1 x 16 tubes: 5.25 seconds 20 µL /well, 384 wells, 1 x 16 tubes: 14 seconds 3 µL /well 1536 wells, 2 x 32 tubes: 7 seconds |
| Volume range |
3 - 3,000 µL/well, selectable in 1 µL increments Minimum Prime Volume: 12 mL |
| Flow rates | User programmable rates from high to low |
| Dispense accuracy |
+1 µL at 5 µL +1 µL at 20 µL +1% at 100 µL |
| Dispense precision |
<5% CV at 5 µL <2.5% CV at 20 µL <1% CV at 100 µL |
| Supply bottle | 1 L or 2 L |
| Sterilization | Chemical, autoclavable option |
Physical Characteristics
| Power | 100 - 240 Volts AC. 50/60 Hz |
| Dimensions | 16.5" W x 18" D x 12.5" H (42 x 46 x 32 cm) |
| Weight | 32 lbs (14.5 kg) |
Regulatory
| Regulatory | CE and TUV marked. ROHS compliant. In Vitro Diagnostic use models are available. |